# Safety checks

```{admonition} TL;DR
:class: tldr

- Each check reads the gradients and RF as they are played and compares them with a limit.
- PNS, mechanical resonance, sound pressure and SAR need scanner data no `.seq` file holds.
- A passing check does not establish that a sequence is safe to run on a scanner or on a subject.
```

An oblique slice puts two logical axes on one gradient coil, and an
echo-planar train can drive the coil at one of its mechanical resonances. You
find out before the scanner does by running {mod}`pypulseqpp.safety` on the
sequence your function returns, with the scanner's limits and models, as in
lesson 7 of the {doc}`Course <../examples/course>`.

## Capabilities

:::{container} capabilities

- **Checks gradient amplitude and slew rate per axis, after each block's rotation.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_max_grad`, {func}`~pypulseqpp.safety.check_max_slew`. Tests: *a gradient stronger than the scanner allows is refused*; *a rotated sequence is weighed on the axes it plays* (`test_safety.py`).

  </details>
- **Reports the largest simultaneous vector amplitude, which a rotation can put on one axis.**

  <details><summary>Show code and tests</summary>

  Code: the report's `vector`. Tests: *the vector peak is what a rotation could put on one axis*; *the vector peak is taken where the axes slew at once* (`test_safety.py`).

  </details>
- **Checks every block boundary for a gradient step, and that every axis ends at zero.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_grad_continuity`. Tests: *a gradient that starts where the last did not end is a jump*; *a sequence that leaves a gradient on has not ramped down* (`test_safety.py`).

  </details>
- **Checks against another scanner's limits, derated or capped.**

  <details><summary>Show code and tests</summary>

  Code: `system=`, {func}`~pypulseqpp.apply_system_derates`, {func}`~pypulseqpp.cap_system`. Test: *a sequence can be weighed against another scanner* (`test_safety.py`).

  </details>
- **Estimates PNS with a chronaxie or SAFE nerve model, over the whole sequence.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_pns`, {class}`~pypulseqpp.safety.ChronaxieModel`, {func}`~pypulseqpp.safety.read_safe_model`. Tests: *a rectangular slew follows the strength-duration curve*; *the SAFE response is upstream's*; *a response over threshold is refused* (`test_pns.py`).

  </details>
- **Turns the gradients to the prescription after each block's own rotation.**

  <details><summary>Show code and tests</summary>

  Code: `rotation=` of {func}`~pypulseqpp.safety.check_pns`, {func}`~pypulseqpp.safety.check_mech_resonance` and {func}`~pypulseqpp.safety.check_spl`; {class}`~pypulseqpp.TransformFOV` for the others. Test: *the prescription is turned after the block's own rotation* (`test_pns.py`, `test_mech_resonance.py`).

  </details>
- **Compares the windowed gradient spectrum with the coil's forbidden bands.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_mech_resonance`, {class}`~pypulseqpp.safety.ForbiddenBand`, {func}`~pypulseqpp.safety.read_forbidden_bands`, {func}`~pypulseqpp.safety.mech_resonance_spectrum`. Tests: *a sustained sinusoid reads its own amplitude*; *a band guards only the axis it names*; *the spectrum reads the amplitude the check reports* (`test_mech_resonance.py`).

  </details>
- **Holds the loudest repetition, played without end, to 140 dB peak and 99 dB(A) average.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_spl`, {func}`~pypulseqpp.safety.read_acoustic_response`. Tests: *the repetition of most gradient energy is evaluated*; *a level above its limit fails the check* (`test_acoustic.py`).

  </details>
- **Estimates local and global SAR per repetition from VOPs, single-channel or pTx.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.check_sar`, {class}`~pypulseqpp.safety.VopModel`, {func}`~pypulseqpp.safety.read_vops`. Tests: *the worst repetition decides*; *SAR goes with the square of the drive*; *the global SAR is the largest over the body models* (`test_sar.py`).

  </details>
- **Applies the VOP file's safety factor to local SAR only.**

  <details><summary>Show code and tests</summary>

  Code: `VopModel.metadata`, `safety_factor=`. Test: *the safety factor multiplies local SAR and its ratio, not global* (`test_sar.py`).

  </details>
- **Expresses SAR as a ratio to a reference pulse, which needs no drive calibration.**

  <details><summary>Show code and tests</summary>

  Code: `reference=` of {func}`~pypulseqpp.safety.check_sar`. Tests: *the ratios do not depend on the drive calibration*; *the global ratio is taken body by body* (`test_sar.py`).

  </details>
- **Reads the VOP files mariepy writes, and MARIE's and MATLAB's layouts.**

  <details><summary>Show code and tests</summary>

  Code: {func}`~pypulseqpp.safety.read_vops`. Tests: *a one-body npz file as mariepy writes it is read*; *a mat file stack is read batch first* (`test_sar.py`).

  </details>
- **Plays the gradients as audio, as MATLAB Pulseq's `sound` does.**

  <details><summary>Show code and tests</summary>

  Code: {meth}`~pypulseqpp.Sequence.sound`. Test: *sound follows MATLAB Pulseq's definition* (`test_sound.py`).

  </details>

:::

## Division of responsibility

- **Its own checks before the scan, and hardware monitoring during it.**
  Pulserver runs the checks on this page on every design, with the limits the
  scanner sends.
- **SAR, RF coil heating and gradient heating, by the vendor's routines.**
  {func}`~pypulseqpp.safety.check_sar` is an estimate under the VOPs and drive
  you supply; on a scanner, VOP ratios are the input the vendor's routine needs
  for pTx.
- **Resampling onto its own rasters.** pypulseqpp checks timing on the rasters
  the sequence declares ({doc}`pulseq-file`).

## Method

```{figure} ../generated/figures/safety_checks_on_one_sequence.png
Three checks on one echo-planar shot, each a played quantity against its
limit: the gradient amplitude against `max_grad`, the nerve response against
its threshold, and a 20 ms window's spectrum against a forbidden band placed
on the train's fundamental.
```

### Per-axis limits and rotation

`max_grad` and `max_slew` are per-axis limits. The axes peak at different
times, so the vector peak is below the root-sum-square of the axis peaks; but
a rotation keeps the vector and moves it between axes. A design with two axes
at `max_grad` at once exceeds it at some oblique prescriptions.

### Nerve response and gradient spectra

A short ramp needs a larger slew rate to stimulate, $S(\tau) =
S_{\mathrm{rh}}(1 + c/\tau)$, and the response depends on the slew history, so
the whole sequence is evaluated from rest. A readout train alternating at echo
spacing $\Delta t$ concentrates its gradient spectrum at $1/(2\,\Delta t)$: for
EPI, the echo spacing is what a forbidden band constrains.

### Reference-pulse SAR ratios

The drive per Hz of $B_1$ is a calibration you rarely have. A pulse the
scanner knows, held at the global limit, fixes it: SAR as a ratio to that
pulse, played under the same VOPs, is independent of the calibration.

The full rules, units and formulas are in
{doc}`../developer-guide/internals/safety-checks`.

## Examples

- {doc}`../generated/gallery/01-course/07_hardware_and_safety_checks`: every
  check on the radial MPRAGE, with its PNS trace and gradient spectrum.
- {doc}`../api/safety`: the functions and their reports.