localizer#
- pulserver.virtual.localizer()[source]#
Return the axial, coronal and sagittal images of a phantom’s proton density, as DICOM.
Each image is the phantom’s proton density in a slab
thicknessthick throughcentre, sampled at the centres ofmatrixbymatrixpixels overfovwith pixelmatrix // 2oncentre, where a reconstruction’s FFT puts it, as the phantom’sproton_densitygives it: the ground truth, with no sequence played. The images carry the geometry of the reconstructed ones, so a console plans on them as it plans on a scanned localizer. Their read and phase directions along the physical axes, which are the patient’s LPS axes for a subject lying head first and supine, are (1, 0, 0) and (0, 1, 0) for the axial plane, (1, 0, 0) and (0, 0, -1) for the coronal one, and (0, 1, 0) and (0, 0, -1) for the sagittal one: the radiological views.- Parameters:
field_t (float) – The magnet’s field, in T, recorded as the imaging frequency.
subject (str, default='') – Patient name of the series.
fov (float, default=0.256) – Field of view of each image, in metres.
matrix (int, default=128) – Pixels along each side.
thickness (float, default=0.005) – Slab thickness, in metres.
centre (Sequence[float], default=(0.0, 0.0, 0.0)) – Physical point the three planes cross at, in metres.