Protocol text blocks#
The text blocks in which the design calls and the interpreter exchange
protocols and RF descriptions (Protocol resolution). The
grammar is implemented in pulserver.protocol for the host and in
pulseg_protocol.h for the interpreter.
RF layout#
The RF a protocol plays depends on the protocol. An evaluation may state it as
an RfLayout, from which a scanner estimates the RF of
a prescription without designing it: the RF definitions of a sequence, the RF
instances of one TR in play order, and for each instance the protocol entry its
amplitude follows.
A definition is one RF pulse: a complex waveform at unit peak magnitude, its
timing and its use. Instances of a definition differ in amplitude and in
frequency and phase offset. In the sequence an evaluation builds, the amplitude
of an instance is relative to the first instance of its definition with a
nonzero amplitude, so a refocusing train with a flip angle schedule is one
definition with one instance per pulse, and RF spoiling, which steps the phase
offset, adds no definition (pypulseqpp.Sequence.rf_instances()).
A control of an instance is the flip angle, or a float user entry, of the plugin’s protocol. It states that the amplitude of the instance is proportional to the entry: a scanner that plays a protocol with another value of the entry multiplies the amplitude by the ratio of that value to the value in the evaluated protocol, which is positive. An instance without a control is played at the amplitude it was evaluated at.
A validation states the amplitude of an instance over the peak_hz the listing
states for its definition, not over the first instance of the definition in the
evaluation of the validation. A protocol that changes a flip angle changes the
peak of the definition in its own evaluation, and the amplitude stated includes
that change, so that the product of the amplitude and the listed peak_hz is the
peak RF amplitude the instance plays.
A layout is one TR: the instances one TR plays, and the TR as its period,
over which a scanner averages their RF. Its size does not depend on the matrix.
The shipped plugins state the first TR of the repetition they design, its shot
repeated once per slice of the largest packet, and a balanced steady state the
TR after its half-angle pulse. A definition is named by its number in the
sequence the evaluation builds, in the order of first play. The listing carries
the definitions of the evaluation at the default protocol, and a layout names
the pulses of the listing only where the protocol does not change which pulses
are played first: the length of a train, a flip angle schedule and the matrix do
not, and a protocol that adds a preparation pulse does. Explicit saturation
bands and a fat saturation are the exception a console needs, since it places
them without listing again: the listing’s evaluation places every band the
plugin offers and its fat saturation, and the shipped plugins number the
saturation pulses after the others, so one placed or removed changes no other
number. A prescan of the chain that repeats at its own TR, such as the
calibration of the EPI plugins, is costed by a scanner over its own
repetition; where it plays more RF power than a TR, the layout is its pulses
over the TR, each numbered as the pulse of the listing it plays, or else the
first of its use.
The layout is an estimate for the prescription. An evaluation that states none states no estimate and is valid, and what the scanner checks before the scan is the stored design (Architecture).
Keys and values#
The keys of a protocol are members of the key enums of
pulserver.protocol, of which ProtocolKey is
the union. A key is a str that equals and hashes as its wire name, so a plain
string indexes a mapping of keys, and a scanner sequence stores a plain string
in protocol as the member it names. A Protocol
maps each key to its value in the units of the sequence function’s argument: seconds for a
time, the argument’s unit for a float, a member of the enum for a
ChoiceParam. The prescription entries bind no
argument and keep the units of the wire.
The wire carries the same protocol as integer microseconds, values in the
entry’s UI unit and, for a choice, the index of the chosen option.
from_wire() and
to_wire() are the only conversions between
the two, and the formatting and parsing functions of
pulserver.protocol are the only conversions between keys and values and
the text of a block. A value block that names no option of a choice is refused
when it is parsed, and a request that names none is invalid when it is given
to validate() as a mapping.
StringListParam(), which builds the enum from option
strings, is deprecated: declare the enum and use ChoiceParam.
Listing and value blocks#
Protocols are exchanged as text blocks delimited by [Protocol] and
[Protocol End]. A listing block carries each entry with its schema, one line
per entry:
TE: int|dropdown|8000|1000|80000|10|us|-2
that is, kind, input mode, current value, minimum, maximum, increment, unit and the dropdown options. The line of a string list is its kind, the index of the chosen option and the options:
imaging_mode: stringlist|1|2d|3d
A value block carries name: value lines only, a string list as the index of
its chosen option.
An option (Options) is an integer user entry, the index of its
choice. Its mode, dropdown with the indices as options (0|1 for a flag),
marks it as an option; an interpreter shows it as a number, like any user entry:
user36_value: int|dropdown|0|0|1|1||0|1
user36_name: description|Navigator: 0 off, 1 on
A list call reads only the option entries of the value block it is given, and
lists the plugin at those options; validate and generate load the plugin at
the options of their request before reading the rest of it against that
listing. An interpreter lists the plugin again when an option entry changes.
The RF blocks follow a reply when the call asks for them, and no block follows
where the evaluation is invalid or states no layout. They are lists of numbers:
each number is ASCII decimal with nine significant digits, the numbers of a list
are separated by single spaces, and each list is one line, so that a reader
takes tokens. [RfDefinitions] follows the listing of a list call asked with
rf_definitions, which evaluates the plugin at its default protocol, with
every explicit saturation band it offers placed, under the scanner limits. Each definition is a definition line, then the sample times,
then the real and the imaginary parts of each channel:
[RfDefinitions]
definition <index> <use> <flip_deg> <peak_hz> <bandwidth_hz> <delay_s> <center_s> <duration_s> <channels> <samples>
<time_s> ...
<real> ...
<imaginary> ...
[RfDefinitions End]
index is the number the runs of a layout name the definition by, from 0 in the
order of first play, and use is the RF use of pypulseqpp. flip_deg and
peak_hz, in degrees and Hz, are those of the first instance with a nonzero
amplitude, and bandwidth_hz is the bandwidth that
pypulseqpp.calc_rf_bandwidth() measures on the sum of the channels, in Hz.
delay_s is the delay of the RF event in its block; center_s, duration_s and
the sample times are in seconds from the start of the event, its delay excluded.
The channels share the sample times. The waveform is scaled to unit peak
magnitude over all channels and samples, with the RF shim applied; the frequency
and phase offsets of the event are playout parameters and are not applied.
[RfLayout] follows the value block of a valid validate reply asked with
rf_layout. The instances are run-length encoded in play order, one run
line for consecutive instances of one definition, control and printed
amplitude, and carry no samples:
[RfLayout]
period <s>
run <index> <amplitude> <control|-> <count>
[RfLayout End]
period is the TR in seconds, over which the instances repeat. The amplitude
of a run is unitless: the peak RF amplitude of its instances over the peak_hz
the listing states for the definition, so that amplitude × peak_hz × waveform
is what an instance plays. The host reads those peaks by evaluating the plugin
at its default protocol as well. A run whose definition the listing does not
state, or states with a peak of zero, carries its amplitude relative to the first
instance of its definition with a nonzero amplitude in the evaluation of the
request, as does every run where the evaluation at the default protocol is
invalid. control is the wire name of the entry the amplitude is proportional
to, or -. A reader skips the blocks it does not know.
Display figures#
An evaluation may state figures of its design that a console shows and the
operator does not set (Display): the echo spacing
of an EPI or FSE train, the receive bandwidth across one pixel of the readout
field of view, and the time one frame of a time series or cine takes. A valid
validate reply that states any ends with a [Display] block, one line per
figure stated:
[Display]
EchoSpacing: <s>
PixelBandwidth: <Hz>
FrameTime: <s>
[Display End]
A console reading no block, from an older server or for a figure not stated, falls back on what it computes itself: the bandwidth over the readout matrix for the pixel bandwidth, the scan time for the frame time. The pixel bandwidth is stated where the readout is oversampled, where that fallback would be too large by the oversampling factor.