Check conventions#
TL;DR
A
.seqfile records the system limits the sequence was designed against, not the system it will be played on. The two differ when the file is used at another site, when a site derates its gradient limits, or when a raster is finer in the design script than on the amplifier.The gradient-derived checks evaluate the physical gradient axes, after each block’s
ROTATIONSextension; the SAR check evaluates the RF waveforms and RF shims. A prescription rotation is passed asrotationtocheck_pns()andcheck_mech_resonance(), and applied withTransformFOVbefore the other gradient checks.Every check returns a boolean verdict and a report, whether or not it passes. Gradient amplitude, slew-rate and continuity reports are in Hz/m and Hz/m/s, mechanical-resonance amplitudes in mT/m, PNS responses as fractions of the model threshold, and SAR in W/kg.
The gradient amplitude, slew-rate and continuity checks read only the sequence and its system limits; the PNS, mechanical-resonance and SAR checks require site or coil data supplied as arguments. Timing is checked separately, by
check_timing().Gradient amplitude, slew rate and continuity are evaluated pointwise, PNS over the whole sequence from rest, and mechanical resonance and SAR over windows. Any window exceeding its threshold or limit makes the verdict false.
The conventions shared by the checks in pypulseqpp.safety: which system
a check evaluates against, the frame it reads gradients in, the units of its
report, and the interval it evaluates over.
Designed and played systems#
A .seq file records the system limits the sequence was designed against,
in its [DEFINITIONS] and in the system its writer used. It does not record
the system it will be played on. The two are identical when the designer
entered the limits of the scanner the sequence will run on, and differ when the
file is used at another site, when a site derates its gradient limits, or when
a raster is finer in the design script than on the amplifier.
Frames, rotations and reports#
The gradient-derived checks — amplitude, slew rate, continuity, PNS and
mechanical resonance — evaluate the physical gradient axes, after each
block’s ROTATIONS extension has been applied. The SAR check is RF-derived: it
evaluates the RF waveforms and RF shims, and no gradient axis or rotation enters
it.
A prescription rotation is a further rotation, composed after each block’s own.
check_pns() and
check_mech_resonance() take one directly, as their
rotation argument. The gradient amplitude, slew-rate and continuity checks
read the rotations the sequence holds, so a prescription is applied to them by
transforming the sequence first with TransformFOV and
checking the result.
Every check returns a boolean verdict and a report. The report is returned
whether or not the check passes, and states the values found, where they were
found and the limit or threshold they were compared with. Gradient amplitude,
slew-rate and continuity reports are in Hz/m and Hz/m/s; gamma on the system
limits converts them to mT/m and T/m/s. Mechanical-resonance amplitudes are in
mT/m, PNS responses are fractions of the model threshold, and SAR is in W/kg.
Required data#
The first three checks read only the sequence and its system limits. The last three require site or coil data that no sequence file contains — a nerve model’s coefficients, a gradient assembly’s forbidden bands, a body model’s VOPs and the transmit chain’s calibration — supplied as arguments.
Timing is checked separately, by
check_timing(), which establishes that every event
time is addressable on the raster its event is played on and that the transmit
and receive dead times are respected. A sequence whose gradients are within
every limit can still be unplayable because one delay is off the raster.
Evaluation interval#
Gradient amplitude, slew rate and continuity are pointwise: every sample of the scan is either within the limit or outside it, and the largest value determines the verdict.
The nerve models are dynamic. Their response at one sample depends on the preceding slew history, so the stimulation check is evaluated over the whole sequence in one pass, starting from rest, rather than over a representative repetition.
The mechanical-resonance and SAR checks are evaluated over windows rather than samples. The resonance check slides a window of a stated length along the sequence and transforms each; the SAR check averages RF energy over each repetition detected from the block definitions. Any window exceeding its threshold or limit makes the verdict false.